Shimadzu QP8000 Mass Spec

 
  • Mass range 10-1500 amu (S model) 10-2000 amu (H model).
  • Rapid data acquisition rates for high throughput analysis (6000 amu per second - 6000 data points per second in centroid mode)
  • API source designed for minimal adduct formation and low thermal fragmentation.
  • Both APCI and ESI probes are included in the standard configuration.
  • APCI flow rates 0.2-2 mL/min.
  • ESI flow rates 0.001- 1 mL/min.
  • Nebulizer gas consumption is 5 L/minute only (maximum). No auxiliary gas is used in desolvation.
  • Mass directed fraction control can be triggered by the total ion chromatogram, multi-ion chromatograms or by mass chromatograms in scan mode. In selected ion monitoring mode the total ion chromatogram triggers the fraction collector.
  • Reliable stable performance over a range of LC operating conditions to maximize sample throughput and performance.
  • Rapid source switching and source maintenance.
  • Supports positive and negative ionization.
  • Small footprint and low cost of ownership.
  • Integrated software control for all devices prevents sample loss in the event of a module failure.
  • Integrated auto-tuning and mass calibration using a built-in syringe controlled by a single click of the mouse.